AK95(95S)英文服务手册-4
Calibrations
If the result from the deflation speed test is
below the limit, check that the cuff hose is
correctly connected to the cuff and that the tube
is not kinked or squeezed.
Otherwise the failure is caused by the BPM
module and it must be replaced and sent to
Gambro for repair.
If the result from the pressure transducer cali-
bration test is outside the limits, it indicates that
the self-calibration function within the BPM
module not is able to adjust the pressure trans-
ducer output within limits. Possible causes are
pressure transducer or electronic circuits mal-
function. The BPM module must then be re-
placed and sent to Gambro for repair.
AK 95 / AK 95 S Service Manual 6:23
HCEN 9138 Rev. 10.2005
Calibrations
This page is intentionally left blank
6:24 AK 95 / AK 95 S Service Manual
HCEN 9138 Rev. 10.2005
Spare Parts List
AK 95 AK 95 S
October 2005
7- Spare Part List
Item Denomination Order no. Remarks
Spare Part List 10.2005
Contents :
Front panel outside 7-1
Front panel inside 7-2
Operators panel 7-3
Blood pump 7-4
Blood pump, Motor unit 7-5
Heparin pump 7-6
Bottom box 7-7
文档评论(0)